Deskripsi
WOHLER A 550 /A 500 L FLUE GAS ANALYZER
PORTABLE FLUE GAS ANALYZER WITH SUPERIOR ACCURACY
No matter how you look at it, the new Wöhler A 550 INDUSTRIAL really is a „nifty piece of equipment“: The individual functions can be launched as intuitively as using a smartphone via the large 7″ color touchscreen. And the brightly lit monitor can be read anywhere. The large screen, the clear arrangement and the graphical presentation of readings allow excellent readability.
Oxygen concentration (O2 ) in flue gas | |||||
Display | Volume % referenced to dry flue gas | ||||
Measurement principle | Electrochemical sensor | ||||
Range | 0.0-21.0 vol. % | ||||
Accuracy | ± 0.3 Vol.-% | ||||
Carbon monoxide (CO 4,000 ppm) in flue gas (Wohler A 550) | |||||
Display | Volume ppm referenced to dry flue gas | ||||
Measurement principle | Electrochemical sensor | ||||
Range | 0-4,000 vol. ppm; resolution 1 vol. ppm | ||||
Accuracy | ±20 ppm (< 400 ppm), otherwise 5% of reading | ||||
Carbon monoxide (CO) in flue gas (Wohler A 550 L) | |||||
Display | Volume ppm referenced to dry flue gas | ||||
Measurement principle | Electrochemical sensor | ||||
Range | 0-35,000 vol. ppm; resolution 1 vol. ppm | ||||
Accuracy | ± 100 ppm (< 1,000 ppm), otherwise 10% of reading (with H2 < 5% of reading) | ||||
Carbon monoxide (COhigh) in flue gas (optional) | |||||
Display | Volume ppm referenced to dry flue gas | ||||
Measurement principle | Electrochemical sensor | ||||
Range | 0-100,000 vol. ppm; resolution 1 vol. ppm | ||||
Accuracy | ± 100 ppm (< 1,000 ppm), otherwise 10% of reading (with H2 < 5 % of reading) | ||||
Nitric oxide concentration (NO) in flue gas (optional) | |||||
Display | Volume ppm referenced to dry flue gas | ||||
Measurement principle | Electrochemical Sensor | ||||
Range | 0-3,000 vol. ppm (continuously up to 1,000); | ||||
Resolution | 1 vol. ppm | ||||
Accuracy | ±5 Vol.-ppm (<100 ppm), otherwise 5 % of reading | ||||
Nitrogen dioxide concentration (NO2 ) in flue gas (optional) | |||||
Display | Volume ppm referenced to dry flue gas | ||||
Measurement principle | Electrochemical sensor | ||||
Range | 0-1000 vol. ppm (continuously up to 200); resolution 1 vol. ppm | ||||
Accuracy | ±5 vol. ppm (< 100 ppm), otherwise 5 % of reading | ||||
Sulfur dioxide (SO2 ) in flue gas (optional) | |||||
Display | Volume ppm referenced to dry flue gas | ||||
Measurement principle | Electrochemical sensor | ||||
Range | 0-5,000 vol. ppm; resolution 1 vol. ppm | ||||
Accuracy | ± 10 vol. ppm (0…200 ppm), otherwise 5 % of reading | ||||
Chimney draft / differential pressure (PD) with 4 Pa test (Wohler A 550) | |||||
Display | inWC | ||||
Measurement principle | Semi-conductor diaphragm | ||||
Range | 0.00 to ± 44.16 inWC; resolution 0.004 inWC (< 4.01 inWC), otherwise 0.004 inWC, with ventilation loss measurement 0.004 inWC | ||||
Accuracy | 0.0012 inWC (< 0.04 inWC), otherwise 3 % of reading Drift < 0.00081 inWC in 5 minutes | ||||
Chimney draft / differential pressure (PD) (Wohler A 550 L) | |||||
Display | inWC | ||||
Measurement principle | Semi-conductor diaphragm | ||||
Range | 0.00 to ± 44.16 inWC; resolution 0.004 inWC | ||||
Accuracy | 0.008 inWC (< 0.16 inWC), otherwise 5 % of reading | ||||
Flue gas temperature (TS) | |||||
Display | °F | ||||
Measurement principle | Thermocouple (NiCr-Ni) (NiCr-Ni) | ||||
Range | -4 °F to 1,472 °F; resolution 32.18 °F | ||||
Accuracy | 32-271.4 °F: ± 35.6°F 271.4-1,472 °F: ± 1.5 % of reading | ||||
Combustion air temperature (TA) | |||||
Display | °F | ||||
Measurement principle | Thermocouple (NiCr-Ni) | ||||
Range | -4 °F to 212 °F; resolution 32.2 °F | ||||
Accuracy | ±33.8°F | ||||
Wood moisture | |||||
Display | Mass of water referenced to dry fuel mass | ||||
Measurement principle | Electrical resistance measurement | ||||
Range | 10.0-40.0%; resolution 0.1% | ||||
Accuracy | ±40% of reading tested to VDI 4206 Part 4 | ||||
Power supply | Lithium-Ion, rechargeable battery 3.7 V, 5800 mAh charges via USB | ||||
Battery operating time | Approx. 12 h (depends on operating status and display illumination) | ||||
Storage temperature | -4 °F to +122 °F | ||||
Operating temperature | +41-104 °F to maintain stated accuracy | ||||
Weight | 2.75 lbs | ||||
Dimensions | 8.7″ x 6.3″ x 2.2″ (without probe) | ||||
Length of cable-hose: | 67 |